In the default configuration, the HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sending application. A remote sender can reassign studies and related workflow records from one patient identity to another. The archive's unmerge does not reverse this.
References
Configurations
No configuration.
History
16 Oct 2026, 15:17
| Type | Values Removed | Values Added |
|---|---|---|
| New CVE |
Information
Published : 2026-10-16 15:17
Updated : 2026-10-16 20:04
NVD link : CVE-2026-105667
Mitre link : CVE-2026-105667
CVE.ORG link : CVE-2026-105667
JSON object : View
Products Affected
No product.
CWE
CWE-306
Missing Authentication for Critical Function