CVE-2021-4479

Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
Configurations

No configuration.

History

No history.

Information

Published : 2026-06-02 20:16

Updated : 2026-07-22 19:10


NVD link : CVE-2021-4479

Mitre link : CVE-2021-4479

CVE.ORG link : CVE-2021-4479


JSON object : View

Products Affected

No product.

CWE
CWE-1286

Improper Validation of Syntactic Correctness of Input