CVE-2019-25723

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Configurations

No configuration.

History

No history.

Information

Published : 2026-06-02 20:16

Updated : 2026-07-22 19:10


NVD link : CVE-2019-25723

Mitre link : CVE-2019-25723

CVE.ORG link : CVE-2019-25723


JSON object : View

Products Affected

No product.

CWE
CWE-1286

Improper Validation of Syntactic Correctness of Input